The standard RTMC can only use LoggerNet as its data source. Currently, RTMC cannot read or display standard web pages within its own display. However, RTMC Pro can accept data fro …
最新内容
Custom Collection exists solely to collect data directly from the data logger and to write those data to one or more data files. Because of this, the data are not stored permanentl …
In RTMC Pro, add the picture or icon to be used as a hyperlink, and then place a hotspot over it. The HotSpot tool is found in the tool box under Commands. …
RTMC is a client to the LoggerNet server. This means that it does not directly interact with the data loggers. Rather, RTMC gets data from the LoggerNet data cache or from collecte …
The TCAV-L is a Type E thermocouple. Use a Type E (chromel –constantan) Thermocouple found in the Sensor | Temperature group. …
Rather than using a running average to find the millivolt output during a calibration, use a single sample with 50 or 60 Hz integration. See Example 1 in the 2014 or later version …
The PWS100 measures the size and velocity of particles in the air. It can provide precipitation rate and water equivalent accumulation, but it is primarily used to report pre …
The potential error is roughly ±0.4% of distance for a ±25° Celsius variation away from 0° Celsius. …
The SR50A-L must be vented to the atmosphere to operate correctly. This is normally not a problem in alpine conditions, but in temperate, damp climates, the vent may introduce cond …
The sensitivity can be adjusted slightly at the factory, but it cannot be recalibrated. Typically, the transducing diaphragm needs to be replaced periodically to maintain calibrati …